Logo del repository
  1. Home
 
Opzioni

An oxygen amperometric gas sensor based on its electrocatalytic reduction in room temperature ionic liquid

TONIOLO, Rosanna
•
DOSSI, Nicolo'
•
SUSMEL, Sabina
altro
DOHERTY AP
2012
  • journal article

Periodico
JOURNAL OF ELECTROANALYTICAL CHEMISTRY
Abstract
A conveniently assembled membrane-free amperometric sensor is proposed for the detection of oxygen in gaseous atmospheres which exploits some profitable properties of room temperature ionic liquids (RTILs), such as their high electrical conductivity, negligible vapour pressure and good thermal stability. The advantages offered by this type of medium were increased by adding small amounts of a further low melting salt bearing a quinone moiety, which allowed the reduction of O-2 to occur through an electrocatalytic pathway taking place at quite lower potentials than those required by its direct reduction. The rate constant (14,160 +/- 370 M-1 s(-1)) of this electrocatalytic process was determined by resorting to linear sweep voltammetric measurements. The performance of this device was assayed under both flowing and static stop-flow conditions on synthetic O-2 + N-2 atmospheres with a controlled oxygen content, which was changed in a wide range (200-10(6) ppm v/v). At room temperature, repeatable (+/- 2.7%) and linearly dependent current signals were recorded, allowing a detection limit of 140 ppm v/v (equivalent to 6.2 x 10(-6) mol of O-2 per L of gaseous atmosphere) to be inferred. At higher temperatures (100 degrees C) a lower detection limit (10 ppm v/v, equivalent to 4.5 x 10(-7) mol of O-2 per L of gaseous atmosphere) could be instead estimated. The possibility of profiting from this sensor for monitoring oxygen under reduced pressure was also assayed, in view of the growing importance attached to the evaluation of residual O-2 in food packaging under vacuum or controlled atmospheres. (C) 2012 Elsevier B.V. All rights reserved.
DOI
10.1016/j.jelechem.2012.02.006
WOS
WOS:000303966700004
Archivio
http://hdl.handle.net/11390/881162
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-84858734448
Diritti
closed access
Soggetti
  • Amperometric oxygen d...

  • Membrane-free amperom...

  • Oxygen monitoring at ...

  • Quinone-mediated elec...

  • Room temperature ioni...

Scopus© citazioni
32
Data di acquisizione
Jun 2, 2022
Vedi dettagli
Web of Science© citazioni
35
Data di acquisizione
Mar 10, 2024
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback