Long-term stability of an injection-molded zirconia bone-level implant: A testing protocol considering aging kinetics and dynamic fatigue

Spies, Benedikt C
•
Maass, Madeleine E.
•
Adolfsson, Erik
altro
Kohal, Ralf Joachim
2017
  • journal article

Periodico
DENTAL MATERIALS
Abstract
Abstract Objective Separately addressing the fatigue resistance (ISO 14801, evaluation of final product) and aging behavior (ISO 13356, standardized sample) of oral implants made from yttria-stabilized zirconia proved to be insufficient in verifying their long-term stability, since (1) implant processing is known to significantly influence transformation kinetics and (2) aging, up from a certain level, is liable to decrease fatigue resistance. Therefore, the aim of this investigation was to apply a new testing protocol considering environmental conditions adequately inducing aging during dynamic fatigue. Methods Zirconia implants were dynamically loaded (107 cycles), hydrothermally aged (85°, 60 days) or subjected to both treatments simultaneously. Subsequent, monoclinic intensity ratios (Xm) were obtained by locally resolved X-ray microdiffraction (μ-XRD2). Transformation propagation was monitored at cross-sections by μ-Raman spectroscopy and scanning electron microscopy (SEM). Finally, implants were statically loaded to fracture. Linear regression models (fracture load) and mixed models (Xm) were used for statistical analyses. Results All treatments resulted in increased fracture load (p ≤ 0.005), indicating the formation of transformation induced compressive stresses around surface defects during all treatment modalities. However, only hydrothermal and combinational treatment were found to increase Xm (p < 0.001). No change in Xm was observed for solely dynamically loaded samples (p ≥ 0.524). Depending on the variable observed, a monoclinic layer thickness of 1–2 μm (SEM) or 6–8 μm (Raman spectroscopy) was measured at surfaces exposed to water during treatments. Significance Hydrothermal aging was successfully induced during dynamic fatigue. Therefore, the presented setup might serve as reference protocol for ensuring pre-clinically long-term reliability of zirconia oral implants.
DOI
10.1016/j.dental.2017.06.002
WOS
WOS:000405360900012
Archivio
http://hdl.handle.net/11368/2906213
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-85020867061
http://www.sciencedirect.com/science/article/pii/S010956411730369X
Diritti
closed access
license:copyright editore
FVG url
https://arts.units.it/request-item?handle=11368/2906213
Soggetti
  • Ceramic

  • Crystallography

  • Dental implant

  • Raman spectroscopy

  • Scanning electron mic...

  • X-ray diffraction

  • Zirconia

  • Materials Science (al...

  • Dentistry (all)

  • Mechanics of Material...

Scopus© citazioni
14
Data di acquisizione
Jun 14, 2022
Vedi dettagli
Web of Science© citazioni
17
Data di acquisizione
Mar 28, 2024
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback