Opzioni
3D-printing for the precision assessment of a new medical device
2022
Periodico
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS. PART C, JOURNAL OF MECHANICAL ENGINEERING SCIENCE
Abstract
Purpose: Additive manufacturing played a keyrole in investigating the precision of a recently-developed device that measures the elastic characteristics of the trabecular bone by simulating the application of loads on a virtual biopsy obtained from radiographic images of the proximal epiphyses in the patient’s hand fingers. The simulation results are combined in a Bone Structure Index (BSI), which has shown to be able to detect trabecular bone alterations due to osteoporosis or other pathological situations. In order to obtain a large number of measurements without having voluntary patients undergo unnecessary radiations, the precision assessment tests were carried out on a 3D-printed phantom hand, in which different mimicked trabecular structures (chips) were inserted. Each mimicked bone had a unique internal structure and density and was 3D-printed using radiopaque composite materials.
Methods: 15 different chips were additively manufactured; 20 measurements were performed on each chip. BSI and BSI_T-score precision values were computed according to ISO 5725 and ISCD standards.
Results: For all the chips, no relationship was found between the mean m_j and standard deviation s_j of the measurements in each chip. The range of the 95% confidence interval (CI) was computed assuming the repeatability standard deviation s_r as the known standard deviation of the measurement method (average of s_j values): 〖CI〗_BSI=±8.14, corresponding to 〖CI〗_Tscore=±0.36. Least Significant Change was evaluated as well: 〖LSC〗_BSI=13.1, corresponding to 〖LSC〗_Tscore=0.58.
Conclusion: The 95% confidence intervals are small when compared to the commonly-accepted diagnostic values, where a patient is classified as osteoporotic if T-score < -2.5, non-osteoporotic if T-score > -1 and osteopoenic if -2.5 < T-score < -1. The LSC results are in line with the requirements for the gold-standard osteoporosis diagnostic systems. Additive manufacturing made it possible to avoid irradiation of humans in this precision assessment.
Diritti
open access
license:copyright editore
license:digital rights management non definito
license uri:iris.pri02
license uri:iris.pri00