Surgical staff doses during liver transplantation on a patient undergoing Lu-177-DOTATATE radioligand therapy

Faustino Bonutti
•
Cinzia Avigo
•
Vittorio Cherchi
altro
Annalisa Drigo
2025
  • journal article

Periodico
RADIATION PROTECTION DOSIMETRY
Abstract
The radiation doses received by the surgical staff during a lifesaving liver transplantation on a female patient, who had received the third cycle of Lu-177-DOTATATE 6 days prior to the surgery, are presented. Thermoluminescent personal dosemeters (TLDs), calibrated in Hp (10) and Hp (0.07), were distributed among the surgical team to assess the effective doses. Finger TLDs, calibrated in Hp (0.07), were specifically provided to the surgeons for hand dose assessment. Ambient dose equivalent rates in air were measured at various distances from the patient, both before and after the liver removal. The results indicate that the overall personal doses were below the minimum detectable dose, with the exception of the first surgeon's finger dose. Additionally, the activity of the liver lesion was measured using a germanium solid-state detector to determine the total activity in the organ at the time of surgery.
DOI
10.1093/rpd/ncaf026
WOS
WOS:001454701800001
Archivio
https://hdl.handle.net/11368/3110178
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-105003598904
https://academic.oup.com/rpd/article/201/6/461/8098731?login=true
Diritti
closed access
license:copyright editore
license uri:iris.pri02
FVG url
https://arts.units.it/request-item?handle=11368/3110178
Soggetti
  • Radiation doses, surg...

  • lifesaving liver tran...

  • Lu-177-DOTATATE

  • radioligand therapy

google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback