Internal breast dosimetry in mammography: Experimental methods and Monte Carlo validation with a monoenergetic x-ray beam

Fedon, Christian
•
Caballo, Marco
•
Longo, Renata
altro
Sechopoulos, Ioannis
2018
  • journal article

Periodico
MEDICAL PHYSICS
Abstract
Purpose: To investigate the performance, such as energy dependence and sensitivity, of thermoluminescent dosimeters (TLD), metal oxide semiconductor field-effect transistor dosimeters (MOSFET), and GafChromicTM films, and to validate the estimates of local dose deposition of a Monte Carlo (MC) simulation for breast dosimetry applications. Methods: Experimental measurements were performed using a monoenergetic beam at the ELETTRA synchrotron radiation light source (Trieste, Italy). The three types of dosimeters were irradiated in a plane transversal to the beam axis and calibrated in terms of air kerma. The sensitivity of MOSFET dosimeters and GafChromicTM films was evaluated in the range of 18–28 keV. Three different calibration curves for the GafChromicTM films were tested (logarithmic, rational, and exponential functions) to evaluate the best-fit curve in the dose range of 1–20 mGy. Internal phantom dose measurements were performed at 20 keV for four different depths (range 0–3 cm, with 1 cm steps) using a homogeneous 50% glandular breast phantom. A GEANT4 MC simulation was modified to match the experimental setup. Thirty sensitive volumes, on the axial-phantom plane were included at each depth in the simulation to characterize the internal dose variation and compare it to the experimental TLD and MOSFET measurements. Experimental 2D dose maps were obtained with the GafChromic TM films and compared to the simulated 2D dose distributions estimated with the MC simulations. Results: The sensitivity of the MOSFET dosimeters and GafChromicTM films increased with x-ray energy, by up to 37% and 48%, respectively. Dose–response curves for the GafChromicTM film result in an uncertainty lower than 5% above 6 mGy, when a logarithmic relationship is used in the dose range of 1–10 mGy. All experimental values fall within the experimental uncertainty and a good agreement (within 5%) is found against the MC simulation. The dose decreased with increasing phantom depth, with the reduction being ~80% after 3 cm. The uncertainty of the empirical measurements makes the experimental values compatible with a flat behavior across the phantom slab for all the investigated depths, while the MC points to a dose profile with a maximum toward the center of the phantom. Conclusions: The calibration procedures and the experimental methodologies proposed lead to good accuracy for internal breast dose estimation. In addition, these procedures can be successfully applied to validate MC codes for breast dosimetry at the local dose level. The agreement among the experimental and MC results not only shows the correctness of the empirical procedures used but also of the simulation parameters.
DOI
10.1002/mp.12792
WOS
WOS:000430182300045
Archivio
http://hdl.handle.net/11368/2925031
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-85045406873
https://aapm.onlinelibrary.wiley.com/doi/abs/10.1002/mp.12792
Diritti
open access
license:creative commons
license uri:http://creativecommons.org/licenses/by/4.0/
FVG url
https://arts.units.it/bitstream/11368/2925031/1/Fedon_et_al-2018-Medical_Physics.pdf
Soggetti
  • breast dosimetry

  • GafChromicTM film

  • GEANT4

  • MOSFET

  • TLD

  • Biophysic

  • Radiology, Nuclear Me...

Web of Science© citazioni
12
Data di acquisizione
Mar 28, 2024
Visualizzazioni
1
Data di acquisizione
Apr 19, 2024
Vedi dettagli
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback