Logo del repository
  1. Home
 
Opzioni

Impact of intravascular enhancement, heart rate, and calcium score on diagnostic accuracy in multislice computed tomography coronary angiography

Cademartiri F.
•
Runza G.
•
Mollet N. R.
altro
Krestin G. P.
2005
  • journal article

Periodico
LA RADIOLOGIA MEDICA
Abstract
PURPOSE: To assess the effect of intravascular enhancement, heart rate, and calcium score on diagnostic accuracy in the detection of significant coronary artery stenosis using 16-row multislice computed tomography (MSCT). MATERIALS AND METHODS: One hundred patients (88 males; 59+/-11 years) with suspected coronary artery disease who had undergone conventional coronary angiography (CA) and MSCT-CA were retrospectively enrolled for the study. Patients underwent a MSCT-CA, with the following protocol: collimation 16x0.75 mm, gantry rotation time 420 ms, feed/rotation 2.8 mm, kV 120, mAs 400-500. The protocol for contrast material administration was 100 ml of Iodixanol 320 mgI/ml at 4 ml/s and the scan delay was defined with a bolus tracking technique. In all patients vascular enhancement was measured in the aortic root, and in the left and right coronary arteries. The average vascular enhancement was used to divide the population in two groups of 50 patients each, one with lower enhancement (Low), and one with higher enhancement (High). In the two groups diagnostic accuracy (per coronary segment) for the detection of significant stenosis (= or >50% lumen reduction) was evaluated in vessels = or >2 mm in diameter using quantitative CA as the reference standard. The differences in diagnostic accuracy were compared with a Chi-square test and a p<0.05 was considered significant. RESULTS: Of the total 1116 segments (= or >2 mm lumen diameter), 173 presented significant stenosis. The sensitivity and specificity for the assessment of significant stenosis were 89.4% and 93.3% vs 94.3% and 97.4% in the presence of increasing intravascular enhancement, 92.8% and 96.7% vs 91.1% and 93.9% in the presence of increasing heart rate, and 89.7% and 97.6% vs 93.3% and 92,8% in the presence of an increasing calcium score. CONCLUSIONS: Increasing intravascular enhancement significantly improves diagnostic accuracy in MSCT-CA. A higher heart rate lowers the specificity in the detection of significant obstructing lesions of the coronary artery. An increasing calcium score determines a lower specificity and a higher sensitivity.
Archivio
http://hdl.handle.net/11368/2297805
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-26244433177
Diritti
metadata only access
Soggetti
  • Aged Artifacts Body W...

  • Statistical Female *H...

  • X-Ray Computed/*metho...

Visualizzazioni
1
Data di acquisizione
Apr 19, 2024
Vedi dettagli
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback