Logo del repository
  1. Home
 
Opzioni

Stroke dynamics of two phaco emulsifiers utilizing new removal modalities

TOGNETTO, DANIELE
2012
  • other

Abstract
Purpose To measure the working frequency and the stroke dynamics of the phaco tip of 3 phacoemulsification machines. Setting University Eye Clinic of Trieste, Italy. Design Experimental study. Methods A video wet fixture was assembled to measure the working frequency using a micro camera and a micropulsed strobe-light system. A different video wet fixture was created to measure tip displacement as vectorial movement at different phaco powers using a microscopic video apparatus. Results The working frequency of the Infiniti Ozil machine was 43.0 kHz in longitudinal mode and 31.6 kHz in torsional mode. The frequency of the Whitestar Signature machine was 29.0 kHz in longitudinal mode and 38.0 kHz with the Ellips FX handpiece. The Stellaris machine had a frequency of 28.8 kHz. The longitudinal stroke of the 3 machines at different phaco powers was statistically significantly different. The Stellaris machine had the highest stroke extent (139 μm). The lateral movement of the Infiniti Ozil and Whitestar Signature machines differed significantly. No movement on the y-axis was observed for the Infiniti Ozil machine in torsional mode. The elliptical path of the Ellips FX handpiece had different x and y components at different phaco powers. Conclusions The 3 phaco machines performed differently in terms of working frequency and stroke dynamics. The knowledge of the peculiar lateral and elliptical path strokes of Infiniti and Whitestar Signature machines may allow the surgeon to fully use these features for lens removal. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned
DOI
10.1016/j.jcrs.2011.11.018
WOS
WOS:000300032200028
Archivio
http://hdl.handle.net/11368/2561640
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-84856836787
Diritti
metadata only access
Soggetti
  • Frequency

  • stroke

  • dynamic

  • phaco emulsifier

  • removal modalities

Scopus© citazioni
0
Data di acquisizione
Jun 14, 2022
Vedi dettagli
Web of Science© citazioni
1
Data di acquisizione
Mar 22, 2024
Visualizzazioni
1
Data di acquisizione
Apr 19, 2024
Vedi dettagli
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback