Logo del repository
  1. Home
 
Opzioni

Bioresorbable vascular scaffold radial expansion and conformation compared to a metallic platform: Insights from in vitro expansion in a coronary artery lesion model

Foin N.
•
Lee R.
•
Bourantas C.
altro
Kimura T.
2016
  • journal article

Periodico
EUROINTERVENTION
Abstract
Aims: This study aimed to compare the acute expansion behaviour of a polymer-based bioresorbable scaffold and a second-generation metallic DES platform in a realistic coronary artery lesion model. Experimental mechanical data with conventional methods have so far shown little difference between metallic stents and currently available polymer-based bioresorbable scaffolds (BRS). Nevertheless, differences in acute results have been observed in clinical studies comparing BRS directly with metallic DES platforms. Methods and results: We examined the expansion behaviour of the bioresorbable vascular scaffold (3.0×18 mm Absorb BVS; Abbott Vascular, Santa Clara, CA, USA) and a metallic DES (3.0×18 mm XIENCE Prime; Abbott Vascular) after expansion at 37°C using identical coronary artery stenosis models (in total 12 experiments were performed). Device expansion was compared during balloon inflation and after deflation using microscopy to allow assessment of plaque recoil. Minimal lumen diameter (MLD) and minimal lumen area (MLA) and stent eccentricity were quantified from optical coherence tomography (OCT) imaging at nominal diameter and after post-dilation at 18 atm. The MLA in the models with BVS deployed was 4.92±0.17 mm2 while in the metallic DES it was 5.40±0.13 mm2 (p=0.02) at nominal pressure (NP), and 5.41±0.20 and 6.07±0.25 mm2 (p=0.02), respectively, after expansion at 18 atm. Stent eccentricity index at the MLA was 0.71±0.02 in BVS compared to 0.81±0.02 in the metal stent at NP (p=0.004), and 0.73±0.03 compared to 0.75±0.02 at 18 atm (p=0.39). Conclusions: Results obtained in this in vitro lesion model were comparable to the results in randomised clinical trials comparing BVS and XIENCE stents in vivo. Such models may be useful in future BRS developments to predict their acute response in vivo in eccentric lesions.
DOI
10.4244/EIJV12I7A138
WOS
WOS:000385692900007
Archivio
http://hdl.handle.net/11368/2962649
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-84990889835
https://eurointervention.pcronline.com/article/bioresorbable-vascular-scaffold-radial-expansion-and-conformation-compared-to-a-metallic-platform-insights-from-in-vitro-expansion-in-a-coronary-artery-lesion-model
Diritti
open access
license:copyright editore
license:digital rights management non definito
FVG url
https://arts.units.it/request-item?handle=11368/2962649
Soggetti
  • Bioresorbable vascula...

  • Conformation

  • Post-dilatation

  • Coronary Disease

  • Finite Element Analys...

  • Human

  • Hydrodynamic

  • Absorbable Implant

  • Blood Vessel Prosthes...

  • Drug-Eluting Stent

  • Tissue Scaffolds

Scopus© citazioni
12
Data di acquisizione
Jun 14, 2022
Vedi dettagli
Web of Science© citazioni
11
Data di acquisizione
Mar 27, 2024
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback