European, Multicenter, Prospective, Non-comparative Clinical Evaluation of an Extended Depth of Focus Intraocular Lens

Auffarth G. U.
•
Moraru O.
•
Munteanu M.
altro
Son H. -S.
2020
  • journal article

Periodico
JOURNAL OF REFRACTIVE SURGERY
Abstract
PURPOSE: To evaluate clinical and safety results obtained with an extended depth of focus (EDOF) intraocular lens (IOL). METHODS: In this European, multicenter, prospective, uncontrolled, interventional study, 77 patients were enrolled in the study, 71 patients received bilateral implantation of the Mini Well Ready EDOF IOL (SIFI S.p.A.), and 68 patients completed the study and were evaluated 2 to 4 months postoperatively. Each clinical examination recorded uncorrected (UDVA) and corrected (CDVA) distance visual acuity, uncorrected (UIVA) and distancecorrected (DCIVA) intermediate visual acuity, and uncorrected (UNVA), distance-corrected (DCNVA), and corrected (CNVA) near visual acuity. A defocus curve from +2.00 to -5.00 diopters (D) was obtained, contrast sensitivity and reading ability were assessed, and the perception of dysphotopsia was evaluated. RESULTS: Mean binocular visual results showed UDVA, UIVA, and UNVA values of -0.01 ± 0.15, 0.03 ± 0.10, and 0.10 ± 0.11 logMAR, respectively. Mean binocular defocus curve demonstrated a visual acuity of 0.30 logMAR or better from +2.00 to -3.00 D. Mean photopic contrast sensitivity values were 1.86, 2.18, 1,97, 1.51, and 1.17 at 1.5, 3.0, 6.0, 12.0, and 18.0 cycles per degree, respectively. A fluent reading speed of 80 words per minute was reached at 0.5 logRAD by 95.31% of patients without distance correction. The mean halo size was 33.06 ± 14.25, mean halo intensity was 38.00 ± 18.51, mean glare size was 23.85 ± 10.43, and mean glare intensity was 42.23 ± 13.22. One postoperative complication, a moderate photophobia, was observed and classified as related to the lens. CONCLUSIONS: The Mini Well Ready EDOF IOL provides good visual acuity across various distances and functional reading ability provided at a near range, and delivers an enhanced contrast sensitivity while causing a low incidence of photic phenomena.
DOI
10.3928/1081597X-20200603-01
WOS
WOS:000565344000001
Archivio
http://hdl.handle.net/11368/2979965
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-85088066628
https://www.healio.com/ophthalmology/journals/jrs/2020-7-36-7/{db7639df-7e25-4bcb-b13b-c3470e391af5}/european-multicenter-prospective-non-comparative-clinical-evaluation-of-an-extended-depth-of-focus-intraocular-lens
Diritti
open access
license:creative commons
license uri:http://creativecommons.org/licenses/by-nc/4.0/
FVG url
https://arts.units.it/bitstream/11368/2979965/1/european-multicenter-prospective-non-comparative-clinical-evaluation-of-an-extended-depth-of-focus-intraocular-lens (1).pdf
Soggetti
  • n/a

Scopus© citazioni
9
Data di acquisizione
Jun 14, 2022
Vedi dettagli
Web of Science© citazioni
15
Data di acquisizione
Mar 27, 2024
Visualizzazioni
1
Data di acquisizione
Apr 19, 2024
Vedi dettagli
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback