The effect of stereotactic body radiotherapy (SBRT) using flattening filter-free beams on cardiac implantable electronic devices (CIEDs) in clinical situations

Aslian, Hossein
•
Kron, Tomas
•
Watts, Troy
altro
Severgnini, Mara
2020
  • journal article

Periodico
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS
Abstract
Purpose This study focused on determining risks from stereotactic radiotherapy using flattening filter‐free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radiosurgery (SRS)/SBRT at the Peter MacCallum Cancer Centre between 2014 and 2018 and b) an experimental study on the impact of FFF beams on CIEDs. Methods A retrospective review was performed. Subsequently, a phantom study was performed using 30 fully functional explanted CIEDs from two different manufacturers. Irradiation was carried out in a slab phantom with 6‐MV and 10‐MV FFF beams. First, a repetition‐rate test (RRT) with a range of beam pulse frequencies was conducted. Then, multifraction SBRT (48 Gy/4 Fx) and single‐fraction SBRT (28 Gy/1 Fx) treatment plans were used for lung tumors delivered to the phantom. Results Between 2014 and 2018, 13 cases were treated with an FFF beam (6 MV, 1400 MU/min or 10 MV, 2400 MU/min), and 15 cases were treated with a flattening filter (FF) beam (6 MV, 600 MU/min). All the devices were positioned outside the treatment field at a distance of more than 5 cm, except for one case, and no failures were reported due to SBRT/SRS. In the phantom rep‐rate tests, inappropriate sensing occurred, starting at a rep‐rate of 1200 MU/min. Cardiac implantable electronic device anomalies during and after delivering VMAT‐SBRT with a 10‐MV FFF beam were observed. Conclusions The study showed that caution should be paid to managing CIED patients when they undergo SBRT using FFF beams, as it is recommended by AAPM TG‐203. Correspondingly, it was found that for FFF beams although there is small risk from dose‐rate effects, delivering high dose of radiation with beam energy greater than 6 MV and high‐dose rate to CIEDs positioned in close vicinity of the PTV may present issues
DOI
10.1002/acm2.12873
WOS
WOS:000544000000015
Archivio
http://hdl.handle.net/11368/2961951
info:eu-repo/semantics/altIdentifier/scopus/2-s2.0-85083167326
https://aapm.onlinelibrary.wiley.com/doi/full/10.1002/acm2.12873
Diritti
open access
license:creative commons
license uri:http://creativecommons.org/licenses/by/4.0/
FVG url
https://arts.units.it/bitstream/11368/2961951/1/acm2.12873.pdf
Soggetti
  • cardiac implantable e...

  • flattening filter‐fre...

  • stereotacticbody radi...

  • volumetric‐modulated ...

Scopus© citazioni
10
Data di acquisizione
Jun 14, 2022
Vedi dettagli
Web of Science© citazioni
13
Data di acquisizione
Mar 26, 2024
Visualizzazioni
1
Data di acquisizione
Apr 19, 2024
Vedi dettagli
google-scholar
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your nstitution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Realizzato con Software DSpace-CRIS - Estensione mantenuta e ottimizzata da 4Science

  • Impostazioni dei cookie
  • Informativa sulla privacy
  • Accordo con l'utente finale
  • Invia il tuo Feedback